微流 HILIC 与离子对 LC−MS 互补扩展单细胞代谢物覆盖,在 BCG 感染巨噬细胞模型中发现旁观者代谢效应
Complementary Single-Cell Microflow HILIC and Ion Pair LC−MS Reveal Bystander Metabolic Effects in a Macrophage Model of Tuberculosis
研究者开发了互补的微流 HILIC 与离子对(IP)LC−MS 方法以扩展单个巨噬细胞的代谢物覆盖,其中离子对 LC−MS 首次用于单细胞分析,两种方法合计得到 633 个带 MS2 谱的命名代谢物。
- 背景
- 单细胞代谢组学受限于单个细胞内代谢物丰度低、化学种类差异大,分析难度高。常规反相条件对高极性和离子型代谢物保留不足,早期流出的这类物质易被漏检。
- 做法
- 作者建立了微量流 HILIC 与离子对(IP)LC−MS 两种互补方法,其中 IP LC−MS 为首次用于单细胞分析。研究对象为 THP-1 巨噬细胞,分 BCG 感染、未感染的旁观者、未暴露对照三组,并将两种方法与既往分析流速 HILIC 方法对比。此外对同一个单细胞依次进行 IP 和 HILIC 分析。
- 结论
- 两种微量流方法检出的特征数均显著多于既往分析流速 HILIC 方法,且覆盖的化学空间互补,合计给出 633 个带 MS2 谱的已命名代谢物。单细胞分辨率的通路分析显示感染相关变化涉及嘌呤、精氨酸、谷胱甘肽和一碳叶酸相关代谢;感染细胞与邻近细胞共享嘌呤和氨基酸变化,旁观者细胞则呈现甘氨酸升高与 ATP 水平异质的潜在表型。
- 意义
- 结果将微量流 HILIC 与 IP LC−MS 定位为互补的正交策略,可在单细胞层面观察复杂微环境中的代谢异质性;同细胞序贯分析为从单个细胞获取最大覆盖提供了可行路径。摘要未给出样本量与效应量,方法在更多体系中的表现仍待验证。
原文摘要
出版方登记 · CrossrefSingle-cell metabolomics remains analytically challenging due to the low abundance and chemical diversity of metabolites in individual cells. We have developed complementary microflow hydrophilic interaction liquid chromatography (HILIC) and ion pair (IP) liquid chromatography-mass spectrometry (LC−MS) methods to expand metabolite coverage in single macrophages. Ion pair LC−MS was applied to single cells for the first time, enabling retention of highly polar and ionic metabolites that elute early under conventional reversed-phase conditions. Across Mycobacterium bovis Bacillus Calmette−Guérin (BCG)-infected, uninfected bystander, and control-unexposed THP-1 macrophages, both microflow methods detected significantly more features than a previously reported analytical-flow HILIC method. The two microflow methods provided complementary chemical space, together yielding 633 unique name-annotated metabolites with MS2 spectra. This depth enabled pathway-level interpretation at single-cell resolution, revealing infection-associated changes in purine-, arginine-, glutathione-, and one-carbon folate-associated metabolism. Metabolite-level interrogation indicated shared purine and amino acid changes in both infected and neighboring macrophages, while a potential bystander phenotype arose, characterized by elevated glycine and heterogeneous adenosine triphosphate (ATP) levels. Finally, we demonstrate sequential IP and HILIC analysis of the same single cell, establishing a route toward maximal coverage from individual cells. These results position microflow HILIC and IP LC−MS as powerful, orthogonal strategies for advancing single-cell metabolomics and unveiling heterogeneity within complex biological microenvironments.
页面未能读取全文,以上为出版方登记的摘要;方法与图表见原文
本页是 AI 精读,不转载论文原文;完整论文(含方法、图表与参考文献)见原文
来源:Analytical Chemistry · doi.org